Technical Specifications
- Particle Size: Total Suspended Particulate Matter (TSPM)
- Flow Rate: 0.6–1.4 m³/min
- Filter Size: 8″ × 10″ GF/A Glass Fibre Filter
- Flow Controller: Digital Flow Controller (DFC)
- Flow Measurement: Digital display with critical orifice
- Timer: Programmable Digital LCD Timer
- Time Totalizer: 0–99.59 hours
- Blower: 3-Phase Brushless Blower
- Motor Control: Variable Frequency Drive (VFD)
- Power Supply: 230V AC, Single Phase, 50 Hz
- Operating Temperature: 0°C to 50°C
- Operating Humidity: 10%–95% RH
- Continuous Operation: At least 28 hours
Key Features
- Automatic flow rate stabilization.
- Electronic compensation for filter loading.
- Automatic shutdown at preset low-flow limit.
- Direct flow measurement using calibrated critical orifice.
- Brushless blower for low-noise operation.
- Provision for gaseous pollutant sampling attachment.
- Weather-resistant gable roof design.
Applications
- Ambient Air Quality Monitoring
- Total Suspended Particulate Matter (TSPM) Sampling
- Industrial Air Pollution Monitoring
- Environmental Monitoring Projects
- Air Quality Research and Laboratory Studies
- Environmental Assessment and Pollution Studies
Product Description
The APM151DFC TSPM Digital High Volume Air Sampler is designed for Total Suspended Particulate Matter (TSPM) monitoring in ambient air. It collects airborne particulate matter on standard 8″ × 10″ Glass Fibre (GF/A) filter papers, making it suitable for environmental monitoring, air quality assessment, research and industrial applications.
Equipped with an advanced Digital Flow Controller (DFC), the APM151DFC automatically maintains a stable sampling flow rate throughout the sampling period. Its electronically controlled flow regulation system compensates for increasing filter resistance caused by dust accumulation, helping maintain consistent sampling conditions.
Technical Operation
The APM151DFC uses an electronically controlled Digital Flow Controller to regulate and stabilize the airflow during particulate sampling.
The system continuously monitors the flow rate and automatically adjusts the blower operation to maintain the selected flow rate. As particulate matter accumulates on the filter, the DFC compensates for the increasing resistance. If the flow rate falls below the preset limit due to excessive filter loading, the controller automatically shuts down the system.
The instrument also features digital flow measurement through a critical orifice, providing real-time flow rate readings in m³/min for convenient monitoring and sampling control.
Key Technical Specifications
- Particle Size : Total Suspended Particulate Matter (TSPM)
- Recommended Filter : GF/A Glass Fibre Filter (8″ × 10″)
- Flow Rate Range : 6–1.4 m³/min
- Flow Control : Electronically controlled Digital Flow Controller (DFC)
- Flow Measurement : Digital display in m³/min using a critical orifice
- Programmable Timer : Digital LCD timer with programmable sampling duration
- Time Totalizer : 0–99.59 hours with digital display and reset provision
- Blower : Heavy-duty, three-phase brushless blower
- Power Supply : 230V AC, single phase, 50 Hz, with ±20V variation, operated through VFD
- Operating Temperature : 0°C to 50°C
- Operating Humidity : 10% to 95%
- Continuous Operation : Designed for continuous operation for at least 28 hours
Digital Flow Controller (DFC) Advantage
The advanced Digital Flow Controller is one of the key features of the APM151DFC.
It automatically regulates the airflow to maintain a consistent sampling rate, even as dust accumulates on the filter. This helps minimize variations in sampling conditions and supports reliable particulate collection.
Key advantages include:
- Automatic flow rate stabilization.
- Electronic compensation for increasing filter resistance.
- User-selectable flow rate between 0.6 and 1.4 m³/min.
- Automatic shutdown when airflow drops below the preset limit.
- Digital flow rate monitoring for easy operation.
Digital Timer and Flow Measurement
The APM151DFC comes with a programmable digital timer that allows users to set the required sampling duration.
The LCD display provides sampling time and flow rate information. Its digital time totalizer records operating hours up to 99.59 hours and includes a reset provision.
The critical orifice-based flow measurement system provides digital flow readings in m³/min, allowing users to monitor the sampling process conveniently.
Heavy-Duty Brushless Blower
The sampler is equipped with a heavy-duty three-phase brushless blower operated through a Variable Frequency Drive (VFD).
Its brushless design eliminates the need for periodic carbon brush replacement associated with conventional brush motors. The system is designed for low-noise operation and continuous sampling for at least 28 hours.
Applications
The APM151DFC High Volume Air Sampler is suitable for:
- Ambient air quality monitoring.
- Total Suspended Particulate Matter (TSPM) sampling.
- Environmental pollution assessment.
- Industrial air quality monitoring.
- Environmental research and laboratory studies.
- Particulate matter collection for environmental analysis.
- Air quality assessment by environmental consultants and monitoring agencies.
Why Choose Serrax APM151DFC?
The Serrax APM151DFC combines automatic flow regulation, digital monitoring and heavy-duty blower technology to support reliable high-volume particulate sampling.
Manufactured by Serrax Technologies, the instrument is backed by technical support and service assistance to meet the requirements of environmental monitoring professionals, laboratories, research institutions and industrial users.
Explore Other Catagories
Air Quality Monitor & Sampler(19)
The Importance of Air Quality Monitoring In a time when environmental issues are increasingly important, the quality of the air we breathe has a direct impact on public health. Numerous health issues and environmental deterioration are caused by air pollution, a global problem with far-reaching effects. In this regard, the importance of air quality monitoring…
Automatic Weather Station(1)
Enhancing Meteorological Precision with Automatic Weather Stations Automatic weather stations have become essential tools for gathering, analyzing, and sharing vital meteorological data in the field of meteorological developments. These high-tech tools have completely changed how climate scientists and meteorologists track and evaluate atmospheric conditions. Automtic weather station provide precise timely information that is essential for…
Food Safety & Inspection(2)
A Comprehensive Exploration of Food Safety and Inspection The safety and quality of the food we eat are, without question, the most important factors in the complex web of our contemporary food chain. The importance of food safety and inspection is highlighted by customers' growing consciousness regarding the source and preparation of their food. An…
Gas Monitors & Analyzers(17)
Essential tools for determining, measuring, and analyzing the concentration of different gases in the air are gas monitors and analyzers. These devices are essential for industrial processes where the presence of certain gases may present risks or affect the quality of the product, environmental monitoring, and worker safety. Now let's take a closer look at…
Laboratory Standard Instruments(11)
Laboratory Standard Instruments are high-accuracy measuring and testing devices designed for research, calibration, and quality assurance in laboratories. These instruments support consistent data, regulatory compliance, and precise analysis across scientific, pharmaceutical, and industrial environments.